Drager Narkomed 2C Anesthesia Machine

Beyond its performance, the Drager Narkomed 2C is a beacon of reliability, thriving in diverse environments. From its temperature resilience to its humidity adaptability, it’s crafted for the rigors of everyday clinical use. With its intuitive monitoring systems, gain insights into oxygen levels, breathing pressures, and more, ensuring you’re always in control. Plus, its seamless data communication, backed by RS-232C compatibility, ensures you’re always connected.

 

Specifications:

General Dimensions:

  • Width x Height x Depth: 32 x 60 1/2 x 27 inches
  • Weight: Approximately 400 lbs

Environmental:

  • Storage Temperature: -20–60°C
  • Humidity (Storage): 10–90% relative humidity (noncondensing)
  • Operating Temperature: 10–40°C
  • Humidity (Operating): 30–75% relative humidity (noncondensing)

Electrical:

  • Equipment Class: IEC 601 Class 1, Type B
  • Leakage Current: ≤ 100 microamps
  • Ground Impedance: ≤ 0.1 ohm (60 Hz source)
  • Dielectric Withstand: ≥ 1500 VAC (per UL 544)
  • Chassis Resistance: ≤ 0.1 ohm (between any metallic point and ground pin on power cord)

117 Volt Power Supply:

  • Primary Input Voltage (Acceptable Range): 90–130 VAC @ 50/60 Hz
  • Primary Input Current: ≤ 6 amps (RMS total), ≤ 1 amp (machine), ≤ 5 amps (receptacles)
  • Primary Input Power: ≤ 780 Watts (includes receptacles)

220/240 Volt Power Supply:

  • Primary Input Voltage (Acceptable Range): 200–260 VAC @ 50/60 Hz
  • Primary Input Current: ≤ 0.4 amp @ 240 VAC, 50 Hz
  • Primary Input Power: ≤ 104 Watts

Backup Battery:

  • Type: Sealed lead-acid, maintenance-free
  • Charging Time: ≤ 16 hours
  • Reserve Power Time: ≥ 30 minutes (from full charge)

Circuit Breakers:

  • Primary AC Power Input: 1.0 amp AC (117 VAC), 0.5 amp AC (240 VAC)
  • Convenience Receptacles: 5.0 amps AC (117 VAC power supply only)
  • Reserve Battery Power: 4.0 amps DC

Miscellaneous:

  • Conductive Caster Resistance: ≤ 250 kohms

Gas Delivery System:

  • Pipeline Inlet Connections: DISS/male, Nut with nipple (Canada)
  • Pipeline Inlet Pressure: 50–55 psi (345–380 kPa) for O2, N2O, and Air
  • Pipeline Gauge Accuracy: ±3 psi (0–25 psi), ±2 psi (25–75 psi), ±3 psi (75–100 psi)
  • Cylinder Connections: Pin-indexed hanger yokes (CGA V-1-1994)
  • Over Pressure Relief Valve: 95 psi (655 kPa), 75 psi (520 kPa) for Canada (CSA Standard Z168.3-M84)
  • Fresh Gas Common Outlet: 15 mm female, 15 mm female and 22 mm male for Canada
  • Fresh Gas Oxygen Concentration (ORC): 25 ±4%
  • Oxygen Flush Flow Rate: 55 (±10) l/min
  • Minimum Oxygen Flow: 150 ±50 ml/min (at 50 psi pipeline pressure)
  • Low Oxygen Supply Pressure Alarm: 34–40 psi
  • Cylinder Gauge Accuracy: ±90 psi (0–750 psi), ±60 psi (750–2250 psi), ±90 psi (2250–3000 psi)

Cylinder Gas Pressures (at 70°F, 21°C):

  • Oxygen, Air: 1900 psi (13100 kPa)
  • Nitrous Oxide: 745 psi (5130 kPa)
  • Carbon Dioxide: 838 psi (5770 kPa)
  • Oxygen-Helium: 1900 psi (13100 kPa)

Flowmeter Accuracy (at 20°C and 760 mmHg):

  • Oxygen, Nitrous Oxide, Air (Fine): 100–1000 ml/min ±2.5% FS
  • Oxygen, Nitrous Oxide, Air (Coarse): 1–10 l/min ±2.5% FS
  • Air (Dual Tapered): 0.2–1 l/min ± 50 ml of reading, 2–10 l/min ±5% FS
  • Carbon Dioxide: 0.05–1.0 l/min ±5% FS
  • Oxygen-Helium: 2–10 l/min ±5% FS
  • Oxygen, Nitrous Oxide (Fine) (Optional, Low-Flow): 20–500 ml/min ±2.5% FS
  • Oxygen, Nitrous Oxide (Coarse) (Optional, Low-Flow): 0.6-10 l/min ±2.5% FS at >1 l/min, ± 15% of reading at <1 l/min
  • Oxygen (Auxiliary Oxygen): 0–10 l/min ±5% FS

Vaporizers (Vapor 19.1):

  • Temperature Range: +15–35°C (at normal atmospheric pressure of 760 mmHg)
  • Flow Range: 0.25-15 l/min
  • Maximum Pressure Load: 150 mmHg (above atmospheric)
  • Maximum Angle of Inclination: 45°
  • Weight: Approximately 7.5 kg
  • Halothane: Adjustment Range 0.2–5 vol%, Accuracy ± 0.15% concentration (volume) or ± 15% (whichever is higher)
  • Enflurane: Adjustment Range 0.3–7 vol%, Accuracy ± 0.2% concentration (volume) or ± 20% (whichever is higher) or + 20%/- 30% with flow settings 6.0–15 l/min and handwheel settings higher than 5.0% volume concentration
  • Isoflurane: Adjustment Range 0.2–5 vol%, Accuracy ± 0.15% concentration (volume) or ± 15% (whichever is higher)
  • Sevoflurane: Adjustment Range 0.3–8 vol%, Accuracy ± 0.2% concentration (volume) or ± 20% (whichever is higher)

Ventilator:

  • Frequency: 1–99, ±1 BPM (in 1 BPM increments)
  • I:E Ratio: Standard range 1:1–1:4.5, ±0.1 (in increments of 0.5); Extended range 4:1, 3:1, 2:1
  • Inspiratory Flow: 10–100 l/min (uncalibrated)
  • Tidal Volume: 50–1500 ml, ±100 ml
  • Pressure Limit Control Adjustment Range: 15–120 cmH20

Absorber System:

  • Inspiratory Valve Mounting Ring Nut Size: M35 x 1
  • Hose Terminal (Inspiratory Valve): 22 mm male
  • Expiratory Valve Mounting Ring Nut Size: M33 x 1
  • Hose Terminal (Expiratory Valve): 22 mm male

PEEP Valve (Optional):

  • Range: Approximately 2–15 cmH2O (continuously adjustable)

Breathing System Pressure Gauge:

  • Range: -20 to +80 cmH2O
  • Smallest Scale Division: 2 cmH2O
  • Nominal Accuracy: -20 to +5 cmH2O (3% FS), +5 to +55 cmH2O (2% FS), +55 to +80 cmH2O (3% FS)
  • Mounting Ring Nut Size: 1 1/8 x 18

APL Valve:

  • Nominal Low Flow Resistance: 2 cmH2O at 8 l/min
  • Hose Terminal: 19 mm male

Breathing Bag Terminal:

  • Bag Terminal: 22 mm male

Oxygen Monitoring:

  • Range: 0–100 vol% O2
  • Resolution: 1 vol% O2
  • Accuracy: ≤ ± 3 vol% O2 (when calibrated within 18 hours, and constant temperature and pressure)
  • Response Time: 25 sec (T90)
  • Zero Drift: ≤ 0.1 vol% O2/month
  • Span Drift: ≤ 1 vol% O2/8 hours
  • Temperature Error: ≤ ± 3% of reading (15° to 40°C)
  • Sensor Service Life: ≥8 months at 25°C, 50% relative humidity, 50% O2 gas mixture (or ≥5000% hour CO2)

Breathing Pressure Monitoring:

  • Numeric Display Range: -10–125 cmH2O
  • Resolution: 1 cmH2O
  • Accuracy: ±3 cmH2O or ±10% of reading, whichever is greater
  • Waveform Display Scales: 0–40, 0–125 cmH2O

Respiratory Volume Monitoring:

  • Minute Volume Display Range: 0.1–99.9 l
  • Resolution: 0.1 l
  • Accuracy: ≤ +10% or 0.1 l, whichever is greater
  • Tidal Volume Display Range: 0.05–9.99 l (Note: the standard bellows will deliver up to 1.5 l)
  • Resolution: 0.01 l
  • Accuracy: ≤ ±10% or 0.01 l, whichever is greater
  • Volume Apnea Threshold: 0.05 l

Respiratory Rate:

  • Numeric Display Range: 2–99
  • Resolution: 1 BPM
  • Accuracy: ≤ ±10% or 1 BPM, whichever is greater

Sensor Flow Range:

  • Flow Range: 5–100 l/minute

Serial Interface:

  • Serial Ports Type: RS-232C, DTE
  • Baud Rate: 300, 2400
  • Parity: Odd, Even, None
  • Data Bits: 7 or 8
  • Stop Bits: 1 or 2
  • Protocols: Vitalink

Goodmedical has an exclusive selection of wholesale deals specifically tailored for resellers who are seeking high-quality refurbished equipment in the US and Mexico. To ensure customer satisfaction and peace of mind, warranty options are offered with our products. If you're interested in learning more about these incredible offers, don't hesitate to get in touch with us right away. Our team is here to provide you with all the necessary information and assist you in making the best choices for your business. Take advantage of this opportunity and unlock a world of possibilities with our reliable and competitively priced equipment. Contact us now to explore the exciting possibilities that await you!